MTS 24 PLACE CENTRIFUGE MTS515060

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-06-08 for MTS 24 PLACE CENTRIFUGE MTS515060 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[670789] The customer reported that the frayed inner green wiring of the retractile cable of the mts centrifuge was exposed. The customer indicated that the issue was discovered during routine yearly maintenance. An exposed or broken wire may have the potential for electrical shock or fire.
Patient Sequence No: 1, Text Type: D, B5


[7875680] The customer was instructed by ocd customer technical services (cts) to discontinue the use of the instrument. The replacement retractile cable part number was provided to the customer. This customer has not logged any complaints against this instrument since this incident. Product labeling cautions the user of the risk of electric shock involved in servicing the instrument. Product labeling provides the user with the proper procedure to follow when repairs are necessary. The root cause for the damaged wire is unk.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2007-00163
MDR Report Key862321
Report Source05,06
Date Received2007-06-08
Date of Report2007-06-08
Date of Event2007-05-15
Date Mfgr Received2007-05-15
Device Manufacturer Date2000-05-01
Date Added to Maude2008-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID HAGEN, SITE MGR, QRC
Manufacturer Street1295 S.W. 29TH AVE.
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239528
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS 24 PLACE CENTRIFUGE
Generic NameCENTRIFUGE
Product CodeKSO
Date Received2007-06-08
Model NumberNA
Catalog NumberMTS515060
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key961212
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-06-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.