MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-05-20 for TOTAL BILIRUBIN 06L45-21 manufactured by Abbott Manufacturing Inc.
[145517250]
Patient identifier: multiple = (b)(6). All available patient information was included. Additional patient details are not available. An evaluation is in process. A follow up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10
[145517251]
The customer reported falsely elevated total bilirubin results were generated for neonates while using the architect (b)(4) analyzer. The following results were provided: sid (b)(6) - architect: 19. 6 mg/dl; alternate method 13. 9 mg/dl (% shift = 41%) the customer further stated that sid (b)(6) generated an architect result of 15. 2 mg/dl, with a alternate method result of 12. 3 mg/dl (% shift = 23. 6%); however, the customer was not sure if the architect result was generated on the (b)(4) analyzer. The customer was utilizing a critical cut off >/= 17 mg/dl. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1628664-2019-00391 |
MDR Report Key | 8623835 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2019-05-20 |
Date of Report | 2019-07-18 |
Date of Event | 2019-03-08 |
Date Mfgr Received | 2019-07-08 |
Device Manufacturer Date | 2019-01-25 |
Date Added to Maude | 2019-05-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT MANUFACTURING INC |
Manufacturer Street | 1921 HURD DRIVE |
Manufacturer City | IRVING TX 75038 |
Manufacturer Country | US |
Manufacturer Postal Code | 75038 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TOTAL BILIRUBIN |
Generic Name | TOTAL BILIRUBIN REAGENT |
Product Code | CIG |
Date Received | 2019-05-20 |
Catalog Number | 06L45-21 |
Lot Number | 54128UQ11 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-05-20 |