TOSOH HLC-723G8 ANALYZER G8 021560

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-05-20 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Corporation.

MAUDE Entry Details

Report Number8031673-2019-00172
MDR Report Key8625972
Report SourceHEALTH PROFESSIONAL
Date Received2019-05-20
Date of Report2019-05-20
Date of Event2019-04-06
Date Mfgr Received2019-04-23
Device Manufacturer Date2017-02-01
Date Added to Maude2019-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DORIA ESQUIVEL
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO 1058623
Manufacturer CountryJA
Manufacturer Postal1058623
Manufacturer G1TOSOH CORPORATION
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, 1058623
Manufacturer CountryJA
Manufacturer Postal Code1058623
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSOH HLC-723G8 ANALYZER G8
Generic NameG8
Product CodeLCP
Date Received2019-05-20
Model NumberG8
Catalog Number021560
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTOSOH CORPORATION
Manufacturer AddressSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, 1058623 JA 1058623


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.