55 - 100 ML CRYOGENIC STORAGE CONTAINER CF-500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-21 for 55 - 100 ML CRYOGENIC STORAGE CONTAINER CF-500 manufactured by Charter Medical, Ltd..

Event Text Entries

[145942138] A single, used device was received for investigation. The returned bag was visually inspected; no issues were observed. The bag was inflated with compressed air and leak tested under water; a leak was detected in the film at the base of the center filler tube port. The leak was verified to be from a hole in the film, approximately 1/32-inch long. No other issues were observed on the returned bag. The user indicated their investigation determined the bag was damaged during handling after it was removed from cryogenic storage and being prepared for thawing. The ifu for this product states "port tubes and film are fragile in the frozen state and breakage may occur. Handle with care. "
Patient Sequence No: 1, Text Type: N, H10


[145942139] The customer reported a bag breakage for a single cf-500 device. The customer indicated their internal investigation determined the bag was damaged during handling at their facility after it was removed from cryogenic storage and being prepared for thawing. The contents of the bag were administered to a patient. The user facility has a broken bag protocol, which requires collection of blood test cultures from the recipient along with prophylactic antibiotics. The blood culture tests were negative. There were no reports of patient injury or death as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2019-00003
MDR Report Key8628777
Date Received2019-05-21
Date of Report2019-05-21
Date of Event2019-04-11
Date Mfgr Received2019-04-11
Device Manufacturer Date2018-07-22
Date Added to Maude2019-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD MEINECKE
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name55 - 100 ML CRYOGENIC STORAGE CONTAINER
Generic Name55 - 100 ML CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2019-05-21
Returned To Mfg2019-04-26
Model NumberCF-500
Catalog NumberCF-500
Lot Number153723
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.