TRANSTAR SINGLE LINE SET MX20984

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-06-07 for TRANSTAR SINGLE LINE SET MX20984 manufactured by Smiths Medical.

Event Text Entries

[669997] The reporter stated that the breakage or leaking from the product. No product was saved. The glue bonding the flushing mechanism to the blue plate transducer becomes loose while attaching a new flush line to the luer lock end below the flush. There was no pt injury or treatment associated with this event.
Patient Sequence No: 1, Text Type: D, B5


[7957198] The customer indicated that there was no product available for return. Since there was no lot number available, the review of the device history record could not be performed. Review of the complaint database for the previous 24 months indicates this is the second reported issue of this type on this product code. The previous event was due to further processing and assembly of the transducer after leaving smiths, not a smiths manufacturing issue. This incident could also be related to the same issue however without product for evaluation this can not be confirmed. Smiths was unable to confirm this issue without benefit of returned product evaluation. An add'l report wille be submitted should info become available that impacts the outcome of smiths' investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616567-2007-00009
MDR Report Key862943
Report Source05,06
Date Received2007-06-07
Date of Report2007-05-09
Date Mfgr Received2007-05-09
Date Added to Maude2008-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTERRIE DAVIS
Manufacturer Street6250 SHIER-RINGS RD.
Manufacturer CityDUBLIN OH 43016
Manufacturer CountryUS
Manufacturer Postal43016
Manufacturer G1MEDEX DE MEXICO, S.A.
Manufacturer StreetCARRETERA MIGUEL ALEMAN,KM21.7
Manufacturer CityAPODACA, MONTERREY
Manufacturer CountryMX
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSTAR SINGLE LINE SET
Generic NamePRESSURE MONITORING SET
Product CodeDPT
Date Received2007-06-07
Model NumberNA
Catalog NumberMX20984
Lot NumberUNK
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key958874
ManufacturerSMITHS MEDICAL
Manufacturer AddressCARLSBAD CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-06-07

© 2024 FDA.report