ELECSYS ANTI-CCP IMMUNOASSAY 05031656190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-05-22 for ELECSYS ANTI-CCP IMMUNOASSAY 05031656190 manufactured by Roche Diagnostics.

Event Text Entries

[145750280] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[145750281] The initial reporter complained of high results for multiple patients tested for elecsys anti-ccp immunoassay (anti-ccp) on a cobas 8000 e 602 module. In (b)(6) 2019 the customer began to use anti-ccp reagent lot 368033 and noticed an abrupt change in the results. The results were reported outside of the laboratory where the rheumatologists complained of a higher than normal "positive" rate and the results did not correspond to the clinical condition for the affected patient. After (b)(6) 2019 the customer noticed that instead of getting "mostly" results of < 8 u/ml and occasional positive results in the 100's u/ml, all anti-ccp results were between 20-60 u/ml. Data for the results for 39 patient samples tested between (b)(6) 2019. None of the patient samples with questionable high anti-ccp results were repeated. There was no allegation that an adverse event occurred. The e602 module serial number was (b)(4). Investigations are ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01906
MDR Report Key8631043
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-05-22
Date of Report2019-09-05
Date of Event2019-01-24
Date Mfgr Received2019-05-01
Date Added to Maude2019-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS ANTI-CCP IMMUNOASSAY
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2019-05-22
Model NumberNA
Catalog Number05031656190
Lot Number368033
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-22

© 2026 FDA.report