DRIVE MEDICAL 10210-2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2007-06-08 for DRIVE MEDICAL 10210-2 manufactured by Medical Depot Inc. (d/b/a Drive Medical Design & Mfg).

Event Text Entries

[661265] User allegedly was in the kitchen, walked down hallway and proceeded to go up approx nine (9) steps when allegedly the leg of the walker allegedly snapped and user tumbled down the stairway. User does not recall the fall. User broke his right hip and fractured his pelvic bone. Family members called 911. User was transported and admitted to hosp in 2007. After treatment and stabilization, user was then transported five days later to rehab. As of three days later, user was still in rehab.
Patient Sequence No: 1, Text Type: D, B5


[7981632] The suspect device appeared to be used for a period of time. The right rear leg is broken off completely at where the lower rivet attached to the side cross bar. The folding/latching mechanism appeared to work properly. We believe that the device broke because the user applied weight to only one leg while walking on steps. Our product instruction has a warning for the user not to use on stairs or escalators.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2438477-2007-00001
MDR Report Key863248
Report Source08
Date Received2007-06-08
Date of Report2007-05-21
Date of Event2007-05-10
Date Mfgr Received2007-05-11
Date Added to Maude2007-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRICH KOLODNY
Manufacturer Street99 SEAVIEW BLVD.
Manufacturer CityPORT WASHINGTON NY 11050
Manufacturer CountryUS
Manufacturer Postal11050
Manufacturer Phone5169984600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRIVE MEDICAL
Generic NameMECHANICAL WALKER
Product CodeNXE
Date Received2007-06-08
Returned To Mfg2007-05-30
Model NumberNA
Catalog Number10210-2
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key847439
ManufacturerMEDICAL DEPOT INC. (D/B/A DRIVE MEDICAL DESIGN & MFG)
Manufacturer AddressPORT WASHINGTON NY US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2007-06-08