MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-05-22 for GLIDE SYSTEM, LAUNDERABLE 3062000000 manufactured by Stryker Medical-kalamazoo.
[145842937]
It was reported that a customer had a glide that was involved in an alleged caregiver shock event. It was reported when the caregiver plugged the power cord into the wall, there were sparks and the caregiver reported feeling an "electrical shock from hand to foot. " it was further reported "the only fault observed with the glide was that the power conductor connection within the actual power plug currently has a disconnect. "
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0001831750-2019-00570 |
MDR Report Key | 8633177 |
Report Source | DISTRIBUTOR |
Date Received | 2019-05-22 |
Date of Report | 2019-05-22 |
Date of Event | 2019-04-03 |
Date Mfgr Received | 2019-04-25 |
Device Manufacturer Date | 2012-06-13 |
Date Added to Maude | 2019-05-22 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. BRIAN THOMPSON |
Manufacturer Street | 3800 EAST CENTRE AVENUE |
Manufacturer City | PORTAGE MI 49002 |
Manufacturer Country | US |
Manufacturer Postal | 49002 |
Manufacturer Phone | 2693292100 |
Manufacturer G1 | STRYKER MEDICAL-KALAMAZOO |
Manufacturer Street | 3800 EAST CENTRE AVENUE |
Manufacturer City | PORTAGE MI 49002 |
Manufacturer Country | US |
Manufacturer Postal Code | 49002 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GLIDE SYSTEM, LAUNDERABLE |
Generic Name | DEVICE, TRANSFER, PATIENT, MANUAL |
Product Code | FMR |
Date Received | 2019-05-22 |
Catalog Number | 3062000000 |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER MEDICAL-KALAMAZOO |
Manufacturer Address | 3800 EAST CENTRE AVENUE PORTAGE MI 49002 US 49002 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-05-22 |