CLINITEK STATUS+ 10379675

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-05-22 for CLINITEK STATUS+ 10379675 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[145914017] The customer stated that repeat testing was performed to confirm correct results which met the clinical picture and a corrected report was issued. There is no reagent or urine available to be returned for investigation. The customer explains that the maintenance may not have been performed as outlined in the operators guide. Proper maintenance and technique were reviewed with the customer. The customer has been advised to perform proper maintenance and has had no further discrepancies.
Patient Sequence No: 1, Text Type: N, H10


[145914018] The customer reported a false negative urine hcg result on the clinitek status+ when compared to a serum quantitative result. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002637618-2019-00062
MDR Report Key8633377
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-05-22
Date of Report2019-05-22
Date of Event2019-04-25
Date Mfgr Received2019-05-01
Date Added to Maude2019-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFELIX AKINRINOLA
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone6107052212
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY, CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINITEK STATUS+
Generic NameCLINITEK STATUS+
Product CodeJIL
Date Received2019-05-22
Catalog Number10379675
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.