CARBOLIME 55-01-0006

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2007-06-08 for CARBOLIME 55-01-0006 manufactured by Allied Healthcare Products, Inc..

Event Text Entries

[670818] It was reported that the carbolime had been a single canister absorber for 1 week. They had done about 10 hours of surgery using sevoflorane at a flow of 1 lpm. The carbolime was a dark color and smelled like something burnt when removed.
Patient Sequence No: 1, Text Type: D, B5


[7975997] A single canister of carbolime is generally good for about 5. 5 hours of surgery. Therefore, at 10 hours of surgery, it was used almost twice as long as it should have been. Carbolime needs to be changed based on the color change as it is being used.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1924066-2007-00006
MDR Report Key863481
Report Source05
Date Received2007-06-08
Date of Report2007-06-08
Date of Event2007-05-26
Date Mfgr Received2007-05-26
Date Added to Maude2008-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street1720 SUBLETTE AVE.
Manufacturer CityST. LOUIS MO 63110
Manufacturer CountryUS
Manufacturer Postal63110
Manufacturer Phone3147712400
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARBOLIME
Generic NameCO2 ABSORBENT
Product CodeCBL
Date Received2007-06-08
Model Number55-01-0006
Lot Number40617
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key958592
ManufacturerALLIED HEALTHCARE PRODUCTS, INC.
Manufacturer AddressST. LOUIS MO 63110 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-06-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.