LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,08 report with the FDA on 2006-09-25 for LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CM manufactured by Bioenterics Corporation.

Event Text Entries

[15037086] Reported as, "leak of access port. "
Patient Sequence No: 1, Text Type: D, B5


[15454829] Visual examination of the returned device determined the access port tubing connector to be a taper i. Analysis of the device noted damage from a sharp instrument to both the band tubing and the port tubing. There was no indication of wear related damage to the portion of the lap-band system returned. Performance tests indicate no leakage at the access port or access port tubing other than damage from sharp instrument noted above. The lab was unable to duplicate or confirm the reported event. There is no other info available at this time from the surgeon to determine the cause of the alleged event. Device labeling addresses the possible outcome of leakage as follows: "deflation of the band may occur due to leakage from the band, the port or the connector tubing. "
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2024601-2006-00563
MDR Report Key863711
Report Source01,05,08
Date Received2006-09-25
Date of Report2006-08-28
Date of Event2006-07-11
Date Mfgr Received2006-08-28
Device Manufacturer Date2000-10-01
Date Added to Maude2007-06-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactTIM KIRKELIE
Manufacturer Street700 WARD DR.
Manufacturer CitySANTA BARBARA CA 93111
Manufacturer CountryUS
Manufacturer Postal93111
Manufacturer Phone8006244261
Manufacturer G1BIOENTERICS CORP.
Manufacturer Street1035 CINDY LANE
Manufacturer CityCARPINTERIA CA 93013
Manufacturer CountryUS
Manufacturer Postal Code93013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 9.75CM
Generic NameDEVICE FOR TREATMENT OF MORBID OBESITY
Product CodeLTT
Date Received2006-09-25
Returned To Mfg2006-09-18
Model NumberNA
Catalog NumberNI
Lot NumberS00LB140
ID NumberNA
Device Expiration Date2002-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagY
Date RemovedV
Device Sequence No1
Device Event Key846068
ManufacturerBIOENTERICS CORPORATION
Manufacturer Address1035 CINDY LANE CARPINTERIA CA 93013 US