GENDEX EXPERT DC 0.822.5214 EXPERT DC65 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2019-05-23 for GENDEX EXPERT DC 0.822.5214 EXPERT DC65 NA manufactured by Gendex Dental Systems.

Event Text Entries

[147340658] An investigation was conducted using interviews and examining three photos of the unit, lag bolts, and wall where it had been attached. It appears that the top bolt came out of the wall and the bottom bolt was bent from the weight of the unit. A root cause could not be determined; however, it appears that the unit had been moved since the initial installation. The re-install may have diminished the wall stud integrity to support the unit. Sales of the expert dc were discontinued on june 24, 2016. This concludes the investigation.
Patient Sequence No: 1, Text Type: N, H10


[147340659] It was reported that the unit fell off the wall. It was reported that unit was caught by the hygienist and hit the patient in the leg, but there was no injury reported for the patient or the operator (hygienist). Unit was reattached to the wall.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2530069-2019-00004
MDR Report Key8638116
Report SourceOTHER
Date Received2019-05-23
Date of Report2019-05-23
Date of Event2019-05-07
Date Mfgr Received2019-05-07
Device Manufacturer Date2010-01-01
Date Added to Maude2019-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ERIKA MARTIN
Manufacturer Street1910 NORTH PENN ROAD
Manufacturer CityHATFIELD PA 19440
Manufacturer CountryUS
Manufacturer Postal19440
Manufacturer G1GENDEX DENTAL SYSTEMS
Manufacturer Street1910 NORTH PENN ROAD
Manufacturer CityHATFIELD PA 19440
Manufacturer CountryUS
Manufacturer Postal Code19440
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGENDEX EXPERT DC
Generic NameEXTRAORAL SOURCE X-RAY SYSTEM
Product CodeEHD
Date Received2019-05-23
Model Number0.822.5214
Catalog NumberEXPERT DC65 NA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGENDEX DENTAL SYSTEMS
Manufacturer Address1910 NORTH PENN ROAD HATFIELD PA 19440 US 19440


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-23