COMMON BILE DUCT EXPLORATION SET N/A C-CDES-100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-05-24 for COMMON BILE DUCT EXPLORATION SET N/A C-CDES-100 manufactured by Cook Inc.

Event Text Entries

[146044285] Initial reporter also sent report to fda: unknown. Pma/510(k) #: not exempt, preamendment. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[146044286] It was reported the basket of a common bile duct exploration set broke during a laparoscopic common bile duct (cbd) procedure. The physician was performing a laparoscopic procedure to retrieve multiple stones from the patient's common bile duct. Upon attempted removal of the stone, the basket broke off of the tip of the catheter. Reinsertion and attempts to retrieve the basket were unsuccessful, as the basket was swept into the duodenum. In additional information received, the patient required an endoscopic retrograde cholangiopancreatography (ercp) for completion of stone and debris removal. A cholangioscopy performed at the time found no evidence of retained basket in the patient. As reported, there have been no adverse effects experienced by the patient due to this occurrence. The device will not be returned for investigation, as it was discarded by the facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2019-01239
MDR Report Key8639649
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-05-24
Date of Report2019-07-01
Date of Event2019-05-05
Date Mfgr Received2019-06-11
Date Added to Maude2019-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMMON BILE DUCT EXPLORATION SET
Generic NameLQR DISLODGER, STONE BILIARY
Product CodeLQR
Date Received2019-05-24
Model NumberN/A
Catalog NumberC-CDES-100
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-24

© 2024 FDA.report