F8 DIALYZER FINISHED ASSY (CASE) 0500164A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-05-24 for F8 DIALYZER FINISHED ASSY (CASE) 0500164A manufactured by Ogden Manufacturing Plant.

Event Text Entries

[146061446] The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.?
Patient Sequence No: 1, Text Type: N, H10


[146061447] A user facility reported that a dialyzer blood leak occurred immediately after the initiation of the patient? S hemodialysis (hd) treatment. There were no alarms from the fresenius 2008k hemodialysis machine used for treatment, as this was an external blood leak that was visually observed leaking from the dialyzer header cap. The patient? S estimated blood loss (ebl) was approximately 3ml. There was no patient injury, adverse events, or medical intervention required as a result of this event. The patient was restarted on a new machine and treatment completed successfully with a new dialyzer. The complaint device was reportedly discarded.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1713747-2019-00166
MDR Report Key8639952
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-05-24
Date of Report2019-07-09
Date of Event2019-05-10
Date Mfgr Received2019-07-05
Device Manufacturer Date2017-02-25
Date Added to Maude2019-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMATTHEW AMARAL
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999758
Manufacturer G1OGDEN MANUFACTURING PLANT
Manufacturer StreetDIRECTOR, SITE QUALITY 475 WEST 13TH STREET
Manufacturer CityOGDEN UT 84404
Manufacturer CountryUS
Manufacturer Postal Code84404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameF8 DIALYZER FINISHED ASSY (CASE)
Generic NameDIALYZER, CAPILLARY, HOLLOW FIBER
Product CodeFJI
Date Received2019-05-24
Catalog Number0500164A
Lot Number17BU04020
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeMO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOGDEN MANUFACTURING PLANT
Manufacturer Address475 WEST 13TH STREET OGDEN UT 84404 US 84404


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-24

© 2024 FDA.report