SJM MASTERS SERIES MECHANICAL HEART VALVE 31MJ-501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2019-05-24 for SJM MASTERS SERIES MECHANICAL HEART VALVE 31MJ-501 manufactured by St. Jude Medical Puerto Rico, Inc..

Event Text Entries

[146091395] An event of improper leaflet function was reported. Morphological and hydrodynamic examination indicated the valve met abbott specifications. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. Hydrodynamic testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally. This test ensures proper leaflet coaptation and hemodynamic performance. The cause of the reported event could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[146091396] On (b)(6) 2019, a 31mm masters valve was implanted in a (b)(6) male. Following implant, an echo revealed improper leaflet function and the patient was reported to be hemodynamically unstable. The valve was exchanged for another 31mm masters valve (serial number: (b)(4)) and the patient is reported to be stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2648612-2019-00035
MDR Report Key8641223
Report SourceFOREIGN
Date Received2019-05-24
Date of Report2019-05-24
Date Mfgr Received2019-04-24
Device Manufacturer Date2018-02-22
Date Added to Maude2019-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL PUERTO RICO, INC.
Manufacturer StreetP.O. BOX 998 LOT 20 B ST.
Manufacturer CityCAGUAS, PUERTO RICO 00725
Manufacturer Postal Code00725
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSJM MASTERS SERIES MECHANICAL HEART VALVE
Generic NameHEART-VALVE, MECHANICAL
Product CodeLWQ
Date Received2019-05-24
Model Number31MJ-501
Catalog Number31MJ-501
Lot Number6332997
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL PUERTO RICO, INC.
Manufacturer AddressP.O. BOX 998 LOT 20 B ST. CAGUAS, PUERTO RICO 00725 00725


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2019-05-24