TRUSYSTEM 7000 DV TS7000DV 1723633

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-28 for TRUSYSTEM 7000 DV TS7000DV 1723633 manufactured by Trumpf Medizin Systeme Gmbh + Co. Kg.

Event Text Entries

[146633114] During inspection, the technician found no problems with the table. However, he discovered damage to the remote's display screen. No repairs or replacements were completed. A review of the table data logs is currently in progress. A follow-up report will be submitted if any new information is discovered.
Patient Sequence No: 1, Text Type: N, H10


[146633115] During a procedure, a trumpf medical trusystem 7000 dv operating table was being articulated into trendelenburg position, when the column lowered on its own. The account reported that there were no injuries to the patient nor impact to the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007143268-2019-00007
MDR Report Key8647416
Date Received2019-05-28
Date of Report2019-05-02
Date of Event2019-05-02
Date Mfgr Received2019-05-02
Device Manufacturer Date2014-07-14
Date Added to Maude2019-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, 07318
Manufacturer CountryGM
Manufacturer Postal07318
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRUSYSTEM 7000 DV
Generic Name7000 DV
Product CodeJEA
Date Received2019-05-28
Model NumberTS7000DV
Catalog Number1723633
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 SAALFELD, 07318 GM 07318


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-28

© 2026 FDA.report