FLEXXICON DIALYSIS CATHETER KIT 11F X 15CM (SHORT-TERM) (DUAL-LUMEN) N/A 5678150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-29 for FLEXXICON DIALYSIS CATHETER KIT 11F X 15CM (SHORT-TERM) (DUAL-LUMEN) N/A 5678150 manufactured by Bard Access Systems.

Event Text Entries

[146648763] The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, applicable manufacture records, and applicable fmea documents. Based on a review of this information, the following was concluded: the complaint of a guidewire stuck within an introducer needle was confirmed and the cause was determined to be use related. The products returned for evaluation were one 0. 035 in. J-tip guidewire in a plastic hoop and one 18 g introducer needle. The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle. The returned product sample was evaluated and the guidewire was confirmed to be broken with the coil wire being unraveled. It appeared that the inner core wire had broken which allowed the outer coil wire surrounding it to unravel. Microscopic examination of the fracture sites revealed the following: narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire. Damage to the inside edge of the introducer needle which can occur if the guidewire is forcefully retracted against the needle, damaging the shape of the sharpened bevel. Biological material was also seen on the wire which may have contributed to the observed guidewire fracture as retraction of the wire together with the biological material could have caused the pieces to become stuck. Normal movement of the guidewire is away from the sharpened bevel and will not damage the wire. If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle. An examination of the wire structure revealed no potential damage/defect related to manufacture of the product. The product ifu cautions:? Do not pull back guidewire over needle bevel as this may sever the end of the guidewire. The introducer needle must be removed first. Also, if unusual resistance is met during manipulation of the guidewire, discontinue the procedure and determine the cause of resistance before proceeding. Withdraw needle and guidewire if cause of resistance cannot be determined.? A lot history review (lhr) of recv2525 showed no other similar product complaint(s) from this lot number.
Patient Sequence No: 1, Text Type: N, H10


[146648764] It was reported that the needle stuck with guide wire. On (b)(6) 2019- returned wire was broken with the coil wire unraveled.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006260740-2019-01490
MDR Report Key8649686
Date Received2019-05-29
Date of Report2019-05-29
Date of Event2019-03-28
Date Mfgr Received2019-05-10
Device Manufacturer Date2018-08-01
Date Added to Maude2019-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKELSEY ERICKSON
Manufacturer Street605 N. 5600 W.
Manufacturer CitySALT LAKE CITY UT 84116
Manufacturer CountryUS
Manufacturer Postal84116
Manufacturer Phone8015225937
Manufacturer G1BARD REYNOSA S.A. DE C.V. -9617592
Manufacturer StreetBLVD. MONTEBELLO #1 PARQUE INDUSTRIAL COLONIAL
Manufacturer CityREYNOSA, TAMAULIPAS
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFLEXXICON DIALYSIS CATHETER KIT 11F X 15CM (SHORT-TERM) (DUAL-LUMEN)
Generic NameCATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Product CodeLFJ
Date Received2019-05-29
Returned To Mfg2019-04-30
Model NumberN/A
Catalog Number5678150
Lot NumberRECV2525
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBARD ACCESS SYSTEMS
Manufacturer Address605 N. 5600 W. SALT LAKE CITY UT 84116 US 84116


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.