T-COAT MICRO HANDLE KERRISON RONGEUR 2130814BT (NL4252-83T)

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-05-30 for T-COAT MICRO HANDLE KERRISON RONGEUR 2130814BT (NL4252-83T) manufactured by S.u.a. Martin Gmbh &co.kg.

Event Text Entries

[146484384] The rongeur was returned without the broken-off tip. Inspection of the device revealed evidence of repair and/or rework by unauthorized third parties: (1) grinding marks were visible in the tip area, diminishing material stability and increasing the risk of breakage; and (2) the screw head had been drilled out, destroying the shank and part of the thread and creating the risk of the remainder of the thread falling out. Review of manufacturing documents showed that the device conformed to specifications at the time of release. Hardness of the returned device was tested and found to conform to specifications.
Patient Sequence No: 1, Text Type: N, H10


[146484385] While in use on a patient during an ortho-spine procedure, the tip of the device broke off and fell into the patient's body. It was retrieved immediately with surgical pickups under direct visualization. The procedure was completed as planned and the patient was discharged the same day.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010386-2019-00001
MDR Report Key8653002
Report SourceDISTRIBUTOR
Date Received2019-05-30
Date of Report2019-05-30
Date of Event2019-03-26
Date Mfgr Received2019-05-01
Device Manufacturer Date2018-12-01
Date Added to Maude2019-05-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MARKUS PAULI
Manufacturer StreetUHLANDSTRASSE 17
Manufacturer CityRIETHEIM-WEILHEIM, 78604
Manufacturer CountryGM
Manufacturer Postal78604
Manufacturer G1S.U.A. MARTIN GMBH &CO.KG
Manufacturer StreetUHLANDSTRASSE 17
Manufacturer CityRIETHEIM-WEILHEIM, 78604
Manufacturer CountryGM
Manufacturer Postal Code78604
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameT-COAT MICRO HANDLE KERRISON RONGEUR
Generic NameMANUAL RONGEUR
Product CodeHAE
Date Received2019-05-30
Catalog Number2130814BT (NL4252-83T)
Lot NumberXMEI05
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerS.U.A. MARTIN GMBH &CO.KG
Manufacturer AddressUHLANDSTRASSE 17 RIETHEIM-WEILHEIM, 78604 GM 78604


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-30

© 2024 FDA.report