Event Text Entries

[146639276] Product event summary: the guidewire was returned and analyzed. The guidewire was unraveled. The distal conductor of the lead was obstructed due to a guidewire stuck in the lumen. Visual analysis of the guidewire indicated damage during use. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[146639297] It was reported that during left ventricular (lv) lead implant the guidewire got stuck after being placed and maneuvered inside the lead, making it hardly removable. The lead was removed and a new lead was implanted. The guidewire was returned, analyzed, and tested out of specification. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220452-2019-00071
MDR Report Key8656042
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-05-30
Date of Report2019-05-30
Date of Event2019-04-15
Date Mfgr Received2019-05-17
Device Manufacturer Date2018-10-25
Date Added to Maude2019-05-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC
Manufacturer Street37A CHERRY HILL DR
Manufacturer CityDANVERS MA 01923
Manufacturer CountryUS
Manufacturer Postal Code01923
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZINGER XT
Generic NameOCCLUDER, CATHETER TIP
Product CodeDQT
Date Received2019-05-30
Returned To Mfg2019-04-30
Model NumberLVZRXT180J
Catalog NumberLVZRXT180J
Lot NumberG18A10971
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC
Manufacturer Address37A CHERRY HILL DR DANVERS MA 01923 US 01923


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-30

© 2024 FDA.report