EV1000 POWER ADAPTER EVPSB220L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-31 for EV1000 POWER ADAPTER EVPSB220L manufactured by Edwards Lifesciences.

Event Text Entries

[146907350] The product is expected to be returned for analysis; however, it has not yet been received. Upon the return of the product a supplemental report will be sent with the investigation results as well as a device history record review. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
Patient Sequence No: 1, Text Type: N, H10


[146907351] As reported, during use of this ev1000 system, there was smoke after the power cord had been plugged in. The user was not sure whether fire or sparks occurred since the power adapter had been kept inside a container, customized by hospital with the purpose of preventing saline/liquid from dropping onto the power supply. Priming of the transducer had not started yet, therefore there was no saline leakage into the adapter at that moment. It was clarified that there was no sound heard either. The nurse quickly unplugged the plug from the socket. After that, the power adapter was replaced with another one from a different ev1000 system. The ev1000m and databox were still functional, only the assembly power adapter was damaged. There was no allegation of patient injury. The device was available for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2015691-2019-01912
MDR Report Key8656948
Date Received2019-05-31
Date of Report2019-05-08
Date of Event2019-05-03
Date Mfgr Received2019-09-17
Date Added to Maude2019-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SAMANTHA EVELEIGH
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9492503939
Manufacturer G1EDWARDS LIFESCIENCES LLC
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEV1000 POWER ADAPTER
Generic NameCOMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION
Product CodeDXG
Date Received2019-05-31
Returned To Mfg2019-08-26
Model NumberEVPSB220L
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES
Manufacturer AddressONE EDWARDS WAY IRVINE CA 92614 US 92614


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.