[146641632] During change of shift intravenous drip handoff and safety check, both rn's noticed the dexmedetomidine bag was completely empty. A new 250ml bag was recently hung by another staff rn on unit. Thirty minutes later, the previous rn and this rn had noticed the bag was already empty and the pump was alarming "air in line. " the settings were checked and confirmed by both rn's. The settings on the pump were all correct, dose of 1. 5 mcg/kg/hr, rate of 28. 13 ml/hr however the whole bag was gone. The volume on the pump still read there was 204ml left. The sigma pump malfunctioned and the patient received 250ml of dexmedetomidine over about a half hour. The patients vital signs were stable and the physician was notified and patient was assessed at the bedside. This rn will continue to monitor patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8658058
MDR Report Key8658058
Date Received2019-05-31
Date of Report2019-05-01
Date of Event2019-04-29
Report Date2019-05-01
Date Reported to FDA2019-05-01
Date Reported to Mfgr2019-05-31
Date Added to Maude2019-05-31
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPECTRUM IQ
Generic NameINFUSION SAFETY MANAGEMENT SOFTWARE
Product CodePHC
Date Received2019-05-31
Model Number3570009
Catalog Number3570009
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER INTERNATIONAL INC.
Manufacturer AddressONE BAXTER PARKWAY DEERFIELD IL 60015 US 60015


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-31

© 2026 FDA.report