AMERIWATER MRO5X AMERIWATER PURIFICATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-30 for AMERIWATER MRO5X AMERIWATER PURIFICATION SYSTEM manufactured by Ameriwater Llc.

Event Text Entries

[146811349] Water purification system for intended use in a dialysis setting has fittings that are made of brass. The brass fittings corrode and leak water. The leakage cause the membranes to lose their prime and system automatic flushing does not get initiated due to the pressure drop. The membranes and the water purification system due to not being flushed automatically stimulate bacterial growth, which becomes high and in an unacceptable numbers. There is potential that this will cause pts infections and pyrogenic reactions. The tests are performed on the purified water using bacterial colony counts in produced water for dialysis and lal (an assay that is used to estimate endotoxins in water) both tests giving numbers that are not acceptable for the product water. Fda safety report id# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5087032
MDR Report Key8658442
Date Received2019-05-30
Date of Report2019-05-28
Date of Event2019-04-29
Date Added to Maude2019-05-31
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAMERIWATER MRO5X AMERIWATER PURIFICATION SYSTEM
Generic NameSUBSYSTEM, WATER PURIFICATION
Product CodeFIP
Date Received2019-05-30
Model NumberMRO5X
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerAMERIWATER LLC
Manufacturer AddressDAYTON OH 45414 US 45414


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-30

© 2026 FDA.report