HILL-ROM ENVELLA BED P0819A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-30 for HILL-ROM ENVELLA BED P0819A manufactured by Hill-rom, Inc..

Event Text Entries

[146890930] The pt had muscle flap surgery on (b)(6) 2019, and the first wound dressing change the complete pt turning was on (b)(6) 2019. When the pt was turned to his side, a pressure injury was identified at the lower back. When wound nurse evaluated the pt, she placed her hand underneath the pt's back and noted that the air bladder for the bed was deflated and the pt's lower back made contact with the bed frame. At that time, hill-rom tech was on the unit and agreed with the wound care nurse and stated that the bed should alarm when there is a deflated bladder. The pt was placed on the unit and agreed with the wound care nurse and stated that the bed should alarm when there is a deflated bladder. The pt was placed on a waffle overlay while the bed was exchanged, which was completed on the same day. Hill-rom support ref number is (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5087041
MDR Report Key8658570
Date Received2019-05-30
Date of Report2019-05-29
Date of Event2019-05-21
Date Added to Maude2019-05-31
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHILL-ROM ENVELLA BED
Generic NameBED, AIR FLUIDIZED
Product CodeINX
Date Received2019-05-30
Returned To Mfg2019-05-21
Model NumberP0819A
Lot Number(21)T117EN1580 # FOUND AFTER U
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2019-05-30

© 2026 FDA.report