MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-31 for EV1000 PLATFORM EV1000A manufactured by Edwards Lifesciences.
[146945919]
The product is expected to be returned for analysis; however, it has not yet been received. Upon the return of the product a supplemental report will be sent with the investigation results. A device history record review was completed and documented that the device met all specifications upon distribution. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
Patient Sequence No: 1, Text Type: N, H10
[146945920]
As reported, there was a problem with the databox connector and the volumeview cable connector. Per follow-up, there was inconsistency in the data displayed and minor physical damage to the databox connector. Another ev1000 was used. The customer could not provide further information. To troubleshoot, they first changed the volumeview cable, but the problem persisted, so they changed out the databox. They then tried to use the previous volumeview cable in the new databox, but it did not work. The platform was available for evaluation.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2015691-2019-01939 |
MDR Report Key | 8659081 |
Date Received | 2019-05-31 |
Date of Report | 2019-04-30 |
Date of Event | 2019-04-30 |
Date Mfgr Received | 2019-07-15 |
Device Manufacturer Date | 2016-01-05 |
Date Added to Maude | 2019-05-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. SAMANTHA EVELEIGH |
Manufacturer Street | ONE EDWARDS WAY |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal | 92614 |
Manufacturer Phone | 9492503939 |
Manufacturer G1 | EDWARDS LIFESCIENCES LLC |
Manufacturer Street | ONE EDWARDS WAY |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal Code | 92614 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | EV1000 PLATFORM |
Generic Name | COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION |
Product Code | DXG |
Date Received | 2019-05-31 |
Model Number | EV1000A |
Catalog Number | EV1000A |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | EDWARDS LIFESCIENCES |
Manufacturer Address | ONE EDWARDS WAY IRVINE CA 92614 US 92614 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-05-31 |