PEDICLE PROBE - CURVED LUMBAR LENKE 07.02072.001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2019-05-31 for PEDICLE PROBE - CURVED LUMBAR LENKE 07.02072.001 manufactured by Zimmer Biomet Spine Inc..

Event Text Entries

[146679118] (b)(4). Current information is insufficient to permit a valid conclusion about the cause of this event. A follow up report will be sent upon completion of the device evaluation.
Patient Sequence No: 1, Text Type: N, H10


[146679119] It was reported that the tip of a lenke probe broke off within the patient's pedicle within surgery and could not be retrieved. The probe was being levered on when it broke. An alternative probe was used to complete the remainder of the procedure.
Patient Sequence No: 1, Text Type: D, B5


[169045036] Additional information: method, results, and conclusions- the returned device was examined. The tip was found to have fractured off likely due to off-axis force. The complaint is confirmed. A review of the manufacturing records did not identify any issues which would have contributed to this event.
Patient Sequence No: 1, Text Type: N, H10


[169045037] It was reported that the tip of a lenke probe broke off within the patient's pedicle within surgery and could not be retrieved. The probe was being levered on when it broke. An alternative probe was used to complete the remainder of the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012447612-2019-00244
MDR Report Key8659204
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2019-05-31
Date of Report2019-11-08
Date of Event2019-05-03
Date Mfgr Received2019-10-11
Device Manufacturer Date2018-08-29
Date Added to Maude2019-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ASHLEY MCPHERSON
Manufacturer Street10225 WESTMOOR DR. NA
Manufacturer CityWESTMINSTER CO 80021
Manufacturer CountryUS
Manufacturer Postal80021
Manufacturer Phone3034437500
Manufacturer G1ZIMMER BIOMET SPINE INC.
Manufacturer Street10225 WESTMOOR DR. NA
Manufacturer CityWESTMINSTER CO 80021
Manufacturer CountryUS
Manufacturer Postal Code80021
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NamePEDICLE PROBE - CURVED LUMBAR LENKE
Generic NamePROBE
Product CodeHXB
Date Received2019-05-31
Returned To Mfg2019-05-16
Model NumberNA
Catalog Number07.02072.001
Lot Number112489
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET SPINE INC.
Manufacturer Address10225 WESTMOOR DR. NA WESTMINSTER CO 80021 US 80021


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-05-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.