T:SLIM X2 INSULIN PUMP 1000354 1000353

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2019-05-31 for T:SLIM X2 INSULIN PUMP 1000354 1000353 manufactured by Tandem Diabetes Care.

Event Text Entries

[146820893] Per the user guide: do not use any other insulin with your pump other than u-100 humalog? Or novolog?. Only humalog? And novolog? Have been tested and found to be compatible for use in the pump. Use of insulin with lesser or greater concentration can result in under delivery or over delivery of insulin. This can cause very high or a very low blood glucose. No product was returned for evaluation. Should new relevant information become available, a supplemental report will be submitted. Device not returned.
Patient Sequence No: 1, Text Type: N, H10


[146820894] It was reported that the customer experienced continuous elevated bg (300 mg/dl) and insulin injections were administered to address bg. Customer alleged pump was not delivering insulin properly. Customer thought pump may have possibly been exposed to moisture or dropped. Pump system check was performed with tandem technical support (cts) and no device issues were identified. Reportedly, admelog was used in the cartridge. Cts informed customer that admelog was not labeled for use with the pump.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3013756811-2019-30063
MDR Report Key8660193
Report SourceCONSUMER
Date Received2019-05-31
Date of Report2019-05-31
Date of Event2018-01-17
Date Mfgr Received2019-05-09
Device Manufacturer Date2017-07-01
Date Added to Maude2019-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJULIA KENSICK
Manufacturer Street11075 ROSELLE ST.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8583666900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameT:SLIM X2 INSULIN PUMP
Generic NameINSULIN PUMP
Product CodeLZG
Date Received2019-05-31
Model Number1000354
Catalog Number1000353
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTANDEM DIABETES CARE