CARBON STEEL SCALPEL BLADES #22 BB522

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-05-31 for CARBON STEEL SCALPEL BLADES #22 BB522 manufactured by Aesculap Ag.

Event Text Entries

[147164820] (b)(4). Investigation: the peel packaging of the blades are not sealed correctly. One side of the peel is open, the anti-corrosion-inlay is not at the right position. Batch history review: the device quality and manufacturing history record have been checked for the available lot number and found to be according to our specifications valid at the time of production. No similar incidents have been filed with products from this batch. Conclusion and root cause: based on the information available as well as a result of our investigation the root cause of the failure is most probably related to a manufacturing error. Rationale: the detected error can be attributed to a production error.
Patient Sequence No: 1, Text Type: N, H10


[147164821] It was reported the packages had insufficient sealing. The reporter indicated that in almost every case the packages of the scalpel blades were completely open. The blades were falling out of the package. This event was noticed prior to surgery, therefore no harm to the patients.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610612-2019-00363
MDR Report Key8660347
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-05-31
Date of Report2019-05-31
Date of Event2019-05-02
Date Facility Aware2019-05-21
Date Mfgr Received2019-05-02
Device Manufacturer Date2018-11-22
Date Added to Maude2019-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSAY CHROMIAK
Manufacturer Street3773 CORPORATE PARKWAY
Manufacturer CityCENTER VALLEY PA 18034
Manufacturer CountryUS
Manufacturer Postal18034
Manufacturer Phone8002581946
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARBON STEEL SCALPEL BLADES #22
Generic NameBASIC INSTRUMENTS
Product CodeGES
Date Received2019-05-31
Returned To Mfg2019-05-15
Model NumberBB522
Catalog NumberBB522
Lot Number4509960181
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age5 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501