COLLECTION SET TUBING PVC (10/PK) 23116

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-31 for COLLECTION SET TUBING PVC (10/PK) 23116 manufactured by Seisa Medical Sales, Llc.

Event Text Entries

[146941659] The device was returned to the service center for evaluation. A visual inspection was performed on the returned device and found the disposable collection set sealed in its original packaging with a long hair inside. In addition, the original equipment manufacturer (oem) reviewed the dhr for the subject device and lot number and no quality issues were noted during the manufacturing process. The oem also reviewed the photo images of the received product and determined that the root cause of the reported event is due to employees exiting and entering the gowning room at the same time, resulting in a hair falling off an employee? S gown. The oem reported that the gowning procedure will be revised to implement turning the gown inside out prior to placing it on the hanger when leaving the clean room, in order to eliminate the possibilities of hair on the product.
Patient Sequence No: 1, Text Type: N, H10


[146941660] The service center was informed that during a therapeutic dilation and curettage (d&c) procedure, the user noted a hair inside the sterile packaging. The device was replaced and the intended procedure was completed. There was no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2019-00896
MDR Report Key8660792
Date Received2019-05-31
Date of Report2019-05-31
Date of Event2019-02-19
Date Mfgr Received2019-02-19
Date Added to Maude2019-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CONNIE TUBERA
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355124
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOLLECTION SET TUBING PVC (10/PK)
Generic NameCOLLECTION SET TUBING PVC
Product CodeHHK
Date Received2019-05-31
Returned To Mfg2019-04-17
Model Number23116
Lot NumberS188620
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSEISA MEDICAL SALES, LLC
Manufacturer Address9005 MONTANA AVENUE EL PASO TX 79925 US 79925


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.