MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-09-29 for MICROSCAN PROMPT INOCULATION SYSTEM-D B1026-10D manufactured by Dade Behring, Inc..
[654466]
Account reported to dade behring that they have observed insufficient growth (no results) on microscan dried overnight gram-positive and gram-negative panel type. Report was associated with only the identified prompt lot. No report of injury or illness associated with this issue.
Patient Sequence No: 1, Text Type: D, B5
[7977210]
Evaluation: performance testing of customer returns and retention samples. Results: in-house testing performed and confirmed insufficient growth issue with customer provided samples and also with retention samples. Conclusion: dade behring has initiated a field correction for this issue and notified customer of the potential for discrepant mic and/or identification results with clinical and qc isolates with the identified prompt inoculation system-d lot.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2919016-2006-00032 |
MDR Report Key | 868114 |
Report Source | 05 |
Date Received | 2006-09-29 |
Date of Report | 2006-09-20 |
Date of Event | 2006-09-20 |
Date Mfgr Received | 2006-09-20 |
Device Manufacturer Date | 2006-05-01 |
Date Added to Maude | 2007-06-21 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | JOSE UNTALAN |
Manufacturer Street | 2040 ENTERPRISE BLVD. |
Manufacturer City | WEST SACRAMENTO CA 95691 |
Manufacturer Country | US |
Manufacturer Postal | 95691 |
Manufacturer Phone | 9163743285 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Removal Correction Number | FCA 000264 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MICROSCAN PROMPT INOCULATION SYSTEM-D |
Generic Name | PANEL INOCULATING SUPPLIES |
Product Code | JSD |
Date Received | 2006-09-29 |
Model Number | NA |
Catalog Number | B1026-10D |
Lot Number | 20071127 |
ID Number | NA |
Device Expiration Date | 2007-11-27 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | NA |
Device Eval'ed by Mfgr | R |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 849649 |
Manufacturer | DADE BEHRING, INC. |
Manufacturer Address | 2040 ENTERPRISE BLVD. WEST SACRAMENTO CA 95691 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2006-09-29 |