[633282] The customer called to report the incubator is reading temps from 36 to 40 degrees. The customer continued to use the incubator despite the out of range temperature (specification is 35 - 39 degrees). The customer claims no samples were compromised.
Patient Sequence No: 1, Text Type: D, B5


[7904658] The customer claimed no erroneous results were reported. The customer was instructed to discontinue use of the incubator and order a replacement.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2006-00178
MDR Report Key870563
Report Source05,06
Date Received2006-09-22
Date of Report2006-09-22
Date of Event2006-08-24
Date Mfgr Received2006-08-24
Date Added to Maude2007-06-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDAVID HAGEN, MGR
Manufacturer Street1295 SW 29TH AVE
Manufacturer CityPOMANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239528
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS INCUBATOR
Generic NameINCUBATOR
Product CodeJTQ
Date Received2006-09-22
Model NumberNA
Catalog NumberMTS9680
Lot Number2250914973
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key851420
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-09-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.