MTS ID TIPMASTER PIPETTOR MTS9640

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-06-27 for MTS ID TIPMASTER PIPETTOR MTS9640 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[632225] The customer reported that mts id tipmaster pipettor dripped fluid. Fluid drip from the pipettor can lead to carry over / cross contamination and incorrect dispense of fluid which may lead to variations in antigen / antibody ratio and possible erroneous test results. The customer observed the issue and took the pipettor out of use, preventing erroneous results from being reported.
Patient Sequence No: 1, Text Type: D, B5


[7964092] The pipette is no longer covered under warranty. The customer was instructed to discontinue use of this instrument. The instrument was not returned for evaluation, and the root cause of this event is unknown.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2007-00182
MDR Report Key870914
Report Source05,06
Date Received2007-06-27
Date of Report2007-06-26
Date of Event2007-05-29
Date Mfgr Received2007-05-29
Device Manufacturer Date2005-06-01
Date Added to Maude2007-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID HAGEN, SITE MANAGER
Manufacturer Street1295 S.W. 29TH AVENUE
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239528
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS ID TIPMASTER PIPETTOR
Generic NameMANUAL PIPETTOR
Product CodeGJG
Date Received2007-06-27
Model NumberNA
Catalog NumberMTS9640
Lot NumberAN
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key854082
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVEVENUE POMPANO BEACH FL 33069 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-06-27