MASK INTERFACE ADAPTOR RT013

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-07-16 for MASK INTERFACE ADAPTOR RT013 manufactured by Fisher & Paykel Healthcare Ltd.

Event Text Entries

[151128646] (b)(4). Method: the complaint rt013 adapter was returned to fisher & paykel healthcare (f&p) regional office in (b)(4) and was inspected by a trained f&p technician. Results: during the inspection, it was found that the tubing was torn on the distal end. Analysis suggests, the tubing of the rt013 was found to be pulled apart near the circuit connector. Conclusion: from this information it appears that the reported issue is user related, rather than an inherent device failure. All rt013 mask interface adaptors are visually inspected for defects prior to leaving production. Those that fail this inspection are rejected. The user instructions which accompany the rt013 contain the following warning: do not crush or stretch tube.
Patient Sequence No: 1, Text Type: N, H10


[151128647] A healthcare facility in (b)(4) reported via a fisher & paykel healthcare (f&p) representative that the rt013 tube interface adaptor was damaged. There was no reported patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611451-2019-00679
MDR Report Key8791994
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-07-16
Date of Report2019-06-16
Date Mfgr Received2019-06-17
Date Added to Maude2019-07-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. FARANAK GOMAROONI
Manufacturer Street173 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9194534000
Manufacturer G1FISHER & PAYKEL HEALTHCARE LTD
Manufacturer Street15 MAURICE PAYKEL PLACE EAST TAMAKI
Manufacturer CityAUCKLAND, 2013
Manufacturer CountryNZ
Manufacturer Postal Code2013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMASK INTERFACE ADAPTOR
Generic NameBZO
Product CodeBZO
Date Received2019-07-16
Model NumberRT013
Catalog NumberRT013
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFISHER & PAYKEL HEALTHCARE LTD
Manufacturer Address15 MAURICE PAYKEL PLACE EAST TAMAKI AUCKLAND, 2013 NZ 2013


Patients

Patient NumberTreatmentOutcomeDate
10 2019-07-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.