ARNDT ENDOBRONCHIAL BLOCKER SET N/A C-AEBS-9.0-78-SPH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-08-02 for ARNDT ENDOBRONCHIAL BLOCKER SET N/A C-AEBS-9.0-78-SPH manufactured by Cook Inc.

Event Text Entries

[153694081] Pma/510(k) #: k160542. This report includes information known at this time. A follow-up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[153694082] It was reported that the balloon on a arndt endobronchial blocker set would not inflate and hold shape during a lung procedure. The bronchial blocker was placed in a single lumen tube from the package. The clinical team placed the balloon inside the lung in a one lung ventilation case. After placement the team attempted to fill the balloon, however, it would not inflate and hold shape. The device was removed from the patient and the balloon was visually inspected. The balloon appeared to be punctured or torn. A different blocker was used successfully in the patient. The arndt endobronchial blocker was not tested prior to being placed in the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2019-01926
MDR Report Key8855770
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-08-02
Date of Report2019-09-24
Date of Event2019-07-11
Date Mfgr Received2019-09-19
Device Manufacturer Date2019-03-18
Date Added to Maude2019-08-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARNDT ENDOBRONCHIAL BLOCKER SET
Generic NameCBI TUBE, TRACHEAL/BRONCHIAL
Product CodeCBI
Date Received2019-08-02
Returned To Mfg2019-07-23
Model NumberN/A
Catalog NumberC-AEBS-9.0-78-SPH
Lot Number9597903
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2019-08-02