[678582] Users mother reported that the battery charger made a loud "popping sound" during the middle of an overnight charge period, and that this was followed by a "burning electrical smell". It was reported that the battery charger is no longer functional, but that the ibot mobility system is fully charged and operating as expected. There was no allegation of harm reported by the user. While no injury was reported as a result of this event, if this event were to recur, there is a potential to cause injury to the user and/or damage to the device. This report corresponds to independence technology.
Patient Sequence No: 1, Text Type: D, B5


[7981358] Note that the ibot mobility system was not returned for eval, only the associated battery charger. The battery charger was evaluated upon return and confirmed to be non-functional. The battery charger was forwarded to engineering for investigation after being returned. Upon review, the battery charger was determined to be inoperative. Internal examination did not reveal any visual or physical evidence of damage to support the reported "burning electrical smell". Engineering could not further troubleshoot the battery charger at this point. The root cause of the reported condition is not yet known. The battery charger was being returned to the mfr for further failure analysis and report. The results of the mfr's investigation are pending receipt. The results of the initial engineering review has been forwarded to the chu for inclusion in complaint records. The mfr's report and any identified further actions will also be forwarded to the chu once complete/known. The user has not reported any recurrence of the described event since the replacement of the battery charger. Please see add'l scanned pages.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003508375-2007-00008
MDR Report Key886245
Report Source04
Date Received2007-07-24
Date of Report2006-11-20
Date of Event2006-10-23
Date Mfgr Received2006-10-23
Device Manufacturer Date2005-12-01
Date Added to Maude2007-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL O'MEARA, DIR
Manufacturer Street45 TECHNOLOGY DR.
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084122200
Manufacturer G1CREATIVE TECHNOLOGY SERVICES
Manufacturer Street7444 HAGGERTY RD
Manufacturer CityCANTON MI 48187
Manufacturer CountryUS
Manufacturer Postal Code48187
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINDEPENDENCE IBOT 4000 MOBILITY SYSTEM
Generic NameSTAIR CLIMBING WHEELCHAIR
Product CodeIMK
Date Received2007-07-24
Returned To Mfg2006-10-26
Model NumberNA
Catalog NumberIT004000
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key864230
ManufacturerINDEPENDENCE TECHNOLOGY, L.L.C.
Manufacturer Address45 TECHNOLOGY DR. WARREN NJ 07059 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-07-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.