ONLINE TDM CARBAMAZEPINE GEN.4 CARB G4 07258062190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-08-07 for ONLINE TDM CARBAMAZEPINE GEN.4 CARB G4 07258062190 manufactured by Roche Diagnostics.

Event Text Entries

[153824434] The investigation believes there is an interferent present affecting the carb4 results from the cobas c501. The patient samples were provided for further investigation. The investigation is currently ongoing. The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[153824435] The initial reporter complained of questionable carb4 carbamazepine results for 1 patient tested on a cobas 6000 c (501) module compared to the carb carbamazepine result from a cobas integra 400 plus. The initial reporter believes the carb4 results are not matching the treatment of the patient. The initial carb4 result from the cobas c501 for a patient under carbamezepine treatment was < 2. 00 ug/ml with a data flag. On (b)(6) 2019 the physician took samples both before and after treatment and the carb4 results from the cobas c501 were < 2. 00 ug/ml with a data flag and < 2. 00 ug/ml with a data flag. The patient samples were sent to another site and tested on a integra 400. The carb results were positive for the samples collected after treatment. The specific results were not provided. The cobas c501 serial number is (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-02868
MDR Report Key8868703
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-08-07
Date of Report2019-08-07
Date of Event2019-06-24
Date Mfgr Received2019-07-12
Date Added to Maude2019-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameONLINE TDM CARBAMAZEPINE GEN.4
Generic NameCARBAMAZEPINE ENZYME IMMUNOASSAY
Product CodeKLT
Date Received2019-08-07
Model NumberCARB G4
Catalog Number07258062190
Lot Number364239
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-08-07