COTTLE MALLET, FLAT NYLON FACE RH1541

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2007-08-02 for COTTLE MALLET, FLAT NYLON FACE RH1541 manufactured by Cardinal Health.

Event Text Entries

[671992] Nylon flakes from mallet fell off into patient during the procedure. Additionally, they retrieved the tip and it was unclear whether any particles got in the surgical site, but they irrigated the site and didn't see anything. The patient was put on antibiotics afterward as a precaution.
Patient Sequence No: 1, Text Type: D, B5


[8129458] One instrument received for evaluation. Upon inspection, it was noted that the plastic replaceable tip was worn. Also noted were scratches and signs of normal wear and tear. A device history review was conducted with no issues noted. It was determined that this device was manufactured in 1999. A historical complaint review was also conducted with no trend found for this type of complaint on this product code. During the device labeling review, it was determined that the labeling states, "note that an examination of all instruments is recommended prior to use in order to decrease the potential for patient injury". Although the exact cause of the damage cannot be determined. It is possible that the damage was due to excessive forces being applied.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423507-2007-00085
MDR Report Key888481
Report Source05
Date Received2007-08-02
Date of Report2007-08-01
Date of Event2007-07-03
Date Mfgr Received2007-07-03
Device Manufacturer Date1999-01-01
Date Added to Maude2007-10-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELE DONATICH
Manufacturer Street1430 WAUKEGAN ROAD, BLDG KB
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8475786412
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOTTLE MALLET, FLAT NYLON FACE
Generic NameMALLET
Product CodeGFJ
Date Received2007-08-02
Returned To Mfg2007-07-16
Model NumberRH1541
Catalog NumberRH1541
Lot NumberH99
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key905383
ManufacturerCARDINAL HEALTH
Manufacturer AddressMCGAW PARK IL 60085 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-08-02

© 2026 FDA.report