VIPER PRIME COMP DRIVER, X25 286750083

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-08-13 for VIPER PRIME COMP DRIVER, X25 286750083 manufactured by Medos International Sàrl Ch.

Event Text Entries

[154129528] (b)(4). The device is available for evaluation. Investigation will be conducted. Follow up will be filed with the investigation results. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate
Patient Sequence No: 1, Text Type: N, H10


[154129529] It was reported that this was a posterior spinal fusion procedure treating spinal stenosis for the lumbar spine on (b)(6) 2019. While the surgeon was using the vprp compressor driver (286750083) with a compressor driver leg and a compressor solid leg, the tip of the compressor driver broke and stuck in a setscrew? S thread hole. The surgeon completed the procedure within a 30-minute delay, leaving the fragment in the patient? S body. The surgeon commented that excessive load was put during the compression. This complaint involves one (1) device.
Patient Sequence No: 1, Text Type: D, B5


[176715031] Product complaint (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1526439-2019-51963
MDR Report Key8887973
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-08-13
Date of Report2019-07-25
Date of Event2018-07-25
Date Mfgr Received2019-10-11
Device Manufacturer Date2016-08-22
Date Added to Maude2019-08-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone6103142063
Manufacturer G1DEPUY SYNTHES SPINE
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal Code02767
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIPER PRIME COMP DRIVER, X25
Generic NameINSTRUMENT, COMPRESSION
Product CodeHWN
Date Received2019-08-13
Returned To Mfg2019-08-16
Model Number286750083
Catalog Number286750083
Lot NumberKM857447
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDOS INTERNATIONAL SàRL CH
Manufacturer AddressCHEMIN-BLANC 38 LE LOCLE 02400 SZ 02400


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-08-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.