IFLOW IFLOW 200 S PROXIMAL FLOW SENSOR 301.328.010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2019-08-16 for IFLOW IFLOW 200 S PROXIMAL FLOW SENSOR 301.328.010 manufactured by Imtmedical Ag.

Event Text Entries

[154626952] Any additional information received from the customer will be included in a follow-up report. Technical support in conjunction with the customer reported alarm 274 - "flow measured proximally is greater than the flow delivered" with 3 pc. Iflow 200 s of batch 4716 even if the circuit system test was successful. So total of affected sensors is 3 pc. Also, they had to switch to bag ventilation while exchanging the proximal flow sensor.
Patient Sequence No: 1, Text Type: N, H10


[154626953] Customer reported alarm 274 - flow measured proximally is greater than the flow delivered. There were 3 pieces sensors affected the iflow 200 s of batch 4716 even if the circuit system test was successful. Customer confirmed that the patient was switch to a manual resuscitation bag while exchanging the proximal flow sensor. There was a patient involvement but no harm and no medical intervention occurred in this reported event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3013407532-2019-00001
MDR Report Key8902818
Report SourceDISTRIBUTOR,FOREIGN
Date Received2019-08-16
Date of Report2018-09-19
Date of Event2018-09-19
Date Mfgr Received2018-09-19
Date Added to Maude2019-08-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STANLEY TAN
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149193324
Manufacturer G1TECHNOCOM SYSTEMS SDN BHD
Manufacturer StreetPLO 1 JALAN FIRMA 1 KAWASAN PERINDUSTRIAN TEBRAU 1
Manufacturer CityJOHOR BAHRU, 81100
Manufacturer CountryMY
Manufacturer Postal Code81100
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIFLOW
Generic NameSET,TUBING AND SUPPORT, VENTILATOR (W/HARNESS)
Product CodeBZO
Date Received2019-08-16
Model NumberIFLOW 200 S PROXIMAL FLOW SENSOR
Catalog Number301.328.010
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerIMTMEDICAL AG
Manufacturer Address10 JALAN KILANG SIME DARBY ENTERPRISE CENTRE#06-06 SINGAPORE, 15941 SN 15941


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-08-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.