MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2006-10-26 for MTS ID TIPMASTER PIPETTOR MTS9640 6900954 manufactured by Micro Typing Systems, Inc..
[667044]
A customer observed that the tipmaster pipettor was not dispensing properly. Inaccurate dispensing of reagents, if undetected, could result in erroneous results. There was no report of pt harm as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
[8129482]
The pipettor was returned to ocd for investigation into the event. The customer's observations were verified. The customer was shipped a replacement pipettor. The root cause of the improper dispensing is unk.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1056600-2006-00226 |
MDR Report Key | 890636 |
Report Source | 05,06 |
Date Received | 2006-10-26 |
Date of Report | 2006-10-26 |
Date of Event | 2005-02-08 |
Date Mfgr Received | 2006-02-08 |
Date Added to Maude | 2007-08-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | DAVID HAGEN, SITE MANAGER |
Manufacturer Street | 1295 S.W. 29TH AVENUE |
Manufacturer City | POMPANO BEACH FL 33069 |
Manufacturer Country | US |
Manufacturer Postal | 33069 |
Manufacturer Phone | 9546239528 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MTS ID TIPMASTER PIPETTOR |
Generic Name | MANUAL PIPETTOR |
Product Code | GJG |
Date Received | 2006-10-26 |
Model Number | MTS9640 |
Catalog Number | 6900954 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 867869 |
Manufacturer | MICRO TYPING SYSTEMS, INC. |
Manufacturer Address | 1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2006-10-26 |