PKG, 1588 AIM CAMERA CONTROL UNIT (CCU) 1588010000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-08-29 for PKG, 1588 AIM CAMERA CONTROL UNIT (CCU) 1588010000 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[156294848] The product was returned for investigation and the reported failure mode was confirmed. Alleged failure: when an image is taken, from the 1588, the preview screen and monitor go black confirmed failure: digital board failure probable root cause: cables connectors digital board main board transition board fiber board intermittence between transition board and ch connector software camera head connected monitors use error manufacturing/service nonconformity the failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10


[156294849] It was reported that there was a loss of image.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2019-00368
MDR Report Key8946209
Date Received2019-08-29
Date of Report2019-08-29
Date of Event2019-08-05
Date Mfgr Received2019-08-05
Device Manufacturer Date2016-04-18
Date Added to Maude2019-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHELLE TO
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePKG, 1588 AIM CAMERA CONTROL UNIT (CCU)
Generic NameCONFOCAL OPTICAL IMAGING
Product CodeOWN
Date Received2019-08-29
Returned To Mfg2019-08-16
Catalog Number1588010000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2019-08-29