Event Text Entries

[6128] Electrosurgical unit alarm sounded when used without grounding pad during surgery. Staff thought it was not necessary to use grounding pad when using machine in bipolar mode. Biomed engineer checked machine and found nothing wrong with unit. According to manual, the unit should be used with grounding pad when in bipolar mode, & thus the unit reacted properly. Out come: user error. No injury to pt. Device not labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Invalid data - on device service/maintenance. Date last serviced: 01-apr-94. Service provided by: user facility biomedical/bioengineering department. Service records available. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, performance tests performed. Results of evaluation: none or unknown, none or unknown, alarm failure. Conclusion: user error caused event. Certainty of device as cause of or contributor to event: invalid data. Corrective actions: inserviced by biomedical engineering dept. Staff. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8965
MDR Report Key8965
Date Received1994-09-23
Date of Report1994-08-12
Date of Event1994-08-02
Date Facility Aware1994-08-02
Report Date1994-08-12
Date Reported to FDA1994-08-12
Date Added to Maude1994-11-02
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Generic NameELECTROSURGICAL CABINET
Product CodeFAR
Date Received1994-09-23
Model NumberSSSSE3B-11AF/E8003
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Implant FlagN
Device Sequence No1
Device Event Key8625
ManufacturerVALLEY LABS, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1994-09-23