X-TRA TRACTION* SHOE COVER XLG 69254

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2007-08-21 for X-TRA TRACTION* SHOE COVER XLG 69254 manufactured by Kunshan Aieda Non-woven Products Co. Ltd..

Event Text Entries

[17020078] A report was received from the user facility that there were several calls regarding falls associated with the use of extratraction shoe covers product. There were no injury or medical intervention. The site expressed concerned of potential liability associated with future incidents. Example given was transport of the transplant patient. Kimberly-clark has no independent knowledge of the event reported above but is relaying the information that was provided by the facility where the incident occurred. This product incident is documented in the kimberly-clark complaint database and identified.
Patient Sequence No: 1, Text Type: D, B5


[17230201] Four potential lot numbers were cited by the user facility. These lots were manufactured from april 2006 though february 2007. The shoe cover is coated with a hot melt skid resistant coating. The contract manufacturer has performed the following analyses on various complaint samples related to skid-resistant shoe cover performance: coefficient of friction; glue application; and traction loss. The manufacture site has reinforced machine maintenance, on-line operator and quality inspectors follow the correct shoe covers manufacturing operation, and attention to appropriate hot melt shoe coating.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9615517-2007-00097
MDR Report Key898872
Report Source06,07
Date Received2007-08-21
Date of Report2007-08-21
Date of Event2007-07-24
Date Mfgr Received2007-07-24
Date Added to Maude2008-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSALLY HERRICK
Manufacturer Street1400 HOLCOMB BRIDGE ROAD
Manufacturer CityROSWELL GA 90076
Manufacturer CountryUS
Manufacturer Postal90076
Manufacturer G1KUNSHAN AIEDA NON-WOVEN PROD.CO. LTD.
Manufacturer Street168 HUAYUAN RD., ZHANGPU TOWN
Manufacturer CityKUNSHAN, JIANGSU PROV. 215321
Manufacturer CountryCH
Manufacturer Postal Code215321
Single Use0
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameX-TRA TRACTION* SHOE COVER XLG
Generic NameSHOE COVER
Product CodeFXP
Date Received2007-08-21
Returned To Mfg2007-07-26
Model NumberNA
Catalog Number69254
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key965381
ManufacturerKUNSHAN AIEDA NON-WOVEN PRODUCTS CO. LTD.
Manufacturer AddressKUNSHAN, JIANGSU PROV. CH 215321


Patients

Patient NumberTreatmentOutcomeDate
10 2007-08-21

© 2026 FDA.report