GE MULTI ABSORBER, DISPOSABLE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-08-14 for GE MULTI ABSORBER, DISPOSABLE manufactured by Ge.

Event Text Entries

[19960349] Item: ge - general electric-multi absorber, disposable. We have had approx 5 of these co2 absorber units arrive with cracks in the plastic, which have caused leakage of anesthetic/o2 gases from the anesthesia machine, in some cases, severe enough to prevent adequate ventilation of the pt. This device is the co2 absorber for use with the ge healthcare s/5 avance anesthesia machine. Although no pts have been harmed as a result of this, the potential for pt harm exists. The cause of the cracks are not known , however, i suspect they occur during the shipping process. Since the plastic case is somewhat thin and flimsy, i suspect that a better protective case may be needed for shipping 6 to the case. The cracks have occurred in different parts of the absorber, and do not appear to be the direct result of a manufacturing defect or design defect. A warning message on the cannister to examine carefully for cracks and test with positive pressure prior to use may be helpful.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5003483
MDR Report Key899935
Date Received2007-08-14
Date of Report2007-08-14
Date of Event2007-08-14
Date Added to Maude2007-09-05
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGE MULTI ABSORBER, DISPOSABLE
Generic NameCO2 ABSORBER
Product CodeBSF
Date Received2007-08-14
ID NumberREF 8003138
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key880782
ManufacturerGE
Manufacturer AddressGREAT BRITAIN


Patients

Patient NumberTreatmentOutcomeDate
101. Deathisabilit 2007-08-14