[182985337] Manufacturer ref# (b)(4). Common device name: nin? Renal balloon-expandable stent. Similar to device under pma/510(k) p100028. Investigation is still in progress.
Patient Sequence No: 1, Text Type: N, H10


[182985338] Description of event according to initial reporter: stent came loose together with the covering protective tube and could no longer be used. Prior to use the protection tube of the stent had been removed. After getting the delivery system to the right place physician wanted to inflate the balloon and expand the stent. By doing this he realised that the stent was not there and the balloon inflated without a stent. Not knowing where the stent had been lost and being afraid it might be somewhere intracorporal physician searched all the vascular way from entry point to stenting area via x-ray. Just by accident a nurse found the stent still sticking in the protection tube which has been put beside. The stent was showing out just a few mm out of the tube. Then a second stent was opened and placed successfully. Patient outcome: the patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002808486-2019-01464
MDR Report Key9004111
Date Received2019-09-12
Date of Report2019-09-12
Date of Event2019-08-20
Date Mfgr Received2019-08-20
Device Manufacturer Date2016-10-25
Date Added to Maude2019-09-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. THOMAS HESSNER KIRK
Manufacturer StreetSANDET 6
Manufacturer CityBJAEVERSKOV DK-4632
Manufacturer CountryDA
Manufacturer PostalDK-4632
Manufacturer Phone56868686
Manufacturer G1WILLIAM COOK EUROPE
Manufacturer StreetSANDET 6
Manufacturer CityBJAEVERSKOV DK-4632
Manufacturer CountryDA
Manufacturer Postal CodeDK-4632
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFORMULA 414 RX BALLOON-EXPANDABLE STENT
Product CodeNIN
Date Received2019-09-12
Returned To Mfg2019-09-03
Catalog NumberFOVX414-14-135-4-16
Lot NumberE3510096
Device Expiration Date2019-10-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerWILLIAM COOK EUROPE
Manufacturer AddressSANDET 6 BJAEVERSKOV 4632 DA 4632


Patients

Patient NumberTreatmentOutcomeDate
10 2019-09-12