TYPE-S HEAD EXTENSION 20-CFHN-SUB2-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2007-09-08 for TYPE-S HEAD EXTENSION 20-CFHN-SUB2-V manufactured by Med-tec Inc. (dba Civco Medical Solutions).

Event Text Entries

[20931622] Civco was notified that a type-s head extension joints unstuck when the patient's body was on the couch at the head extension when the patient fell and was injured on the elbow. Patient discomfort but no serious outcome was attributed to adverse event.
Patient Sequence No: 1, Text Type: D, B5


[21063887] The device was designed with epoxy holding together the supports for the device. The strength of the epoxy can decrease when subjected to impact loads caused by collisions or improper handling. The device had impact markers possible from previous improper handling. Over time the strength can decrease enough for failure to occur. Additional information about use of product during failure or details about injuries were not known by distributor. Patient information was not provided by the distributor. Instructions for use describes and diagrams correct repositioning process of patient/clinician. Instructions for use indicates to inspect for damage/wear prior to use and not to use a damaged device. Adverse event occurred in another country. Additional contact information was not available.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932738-2007-00002
MDR Report Key910968
Report Source08
Date Received2007-09-08
Date of Report2007-08-09
Date Mfgr Received2007-08-22
Device Manufacturer Date2003-02-25
Date Added to Maude2007-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street1401 8TH STREET SOUTHEAST
Manufacturer CityORANGE CITY IA 51021
Manufacturer CountryUS
Manufacturer Postal51021
Manufacturer Phone7127378688
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTYPE-S HEAD EXTENSION
Generic NameCARBON FIBER COUCH INSERT
Product CodeJAI
Date Received2007-09-08
Returned To Mfg2007-08-22
Model Number20-CFHN-SUB2-V
Catalog Number20-CFHN-SUB2-V
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key897527
ManufacturerMED-TEC INC. (DBA CIVCO MEDICAL SOLUTIONS)
Manufacturer Address1401 8TH STREET SOUTHEAST ORANGE CITY IA 51041 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-09-08

© 2024 FDA.report