BENCHMARK ULTRA STAINER MODULE 05342716001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-09-25 for BENCHMARK ULTRA STAINER MODULE 05342716001 manufactured by Ventana Medical Systems.

Event Text Entries

[165643668] During the visit from the local field service engineer, it was indicated that the air filter tubing going to the waste manifold fitting broke, causing fluid to pour onto the power plug causing to short. The customer site did not want the instrument replaced; as such, parts were removed and exchanged. After the replacement of parts, the local field service engineer performed an operational qualification run on the instrument. The instrument is now performing within acceptable operating specifications. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[165643669] A us customer stated that there was smoke coming from the benchmark ultra stainer module, and there was no alarm or error generated. The customer indicated that the instrument automatically shut off. The operator left the lab temporarily because of the quantity of smoke. There was no harm or injury indicated, and lab operator did not seek any medical attention. A local field service engineer was dispatched to the customer site.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028492-2019-00019
MDR Report Key9120312
Date Received2019-09-25
Date of Report2019-09-25
Date of Event2019-08-30
Date Mfgr Received2019-08-30
Date Added to Maude2019-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA STACIE-ANN CREIGHTON
Manufacturer Street1910 E INNOVATION PARK DR NA
Manufacturer CityTUSCON AZ 85755
Manufacturer CountryUS
Manufacturer Postal85755
Manufacturer Phone9082537112
Manufacturer G1NA
Manufacturer StreetNA NA
Manufacturer CityNA
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameBENCHMARK ULTRA STAINER MODULE
Generic NameSLIDE STAINER, AUTOMATED
Product CodeKPA
Date Received2019-09-25
Model NumberNA
Catalog Number05342716001
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVENTANA MEDICAL SYSTEMS
Manufacturer Address1910 E. INNOVATION PARK DR. NA TUCSON AZ 85755 US 85755


Patients

Patient NumberTreatmentOutcomeDate
10 2019-09-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.