MTS ID TIPMASTER PIPETTOR MTS9640

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-09-17 for MTS ID TIPMASTER PIPETTOR MTS9640 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[15560943] The customer reported that they had observed an incorrect amount of fluid dispensed and aspirated from the mts id tipmaster pipettor. Incorrect dispense of fluid may lead to variations in antigen/antibody ratio and possible erroneous test results. The customer observed the issue and took the pipettor out of use, preventing erroneous results from being reported.
Patient Sequence No: 1, Text Type: D, B5


[15849685] No definitive root cause was determined. The customer indicated that cleaning, re-greasing and replacing the o-rings on the pipettor has resolved the issue. The pipettor passed the volume verification testing and it's operating as expected. This customer has not logged any complaints against this pipettor since this incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2007-00270
MDR Report Key914193
Report Source05,06
Date Received2007-09-17
Date of Report2007-09-17
Date of Event2007-08-23
Date Mfgr Received2007-08-23
Device Manufacturer Date2006-04-01
Date Added to Maude2007-10-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationUNKNOWN
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactAWILDA GERBER, PRINCPL SCIENTIST
Manufacturer Street1295 S.W. 29TH AVE.
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239567
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS ID TIPMASTER PIPETTOR
Generic NameMANUAL PIPETTOR
Product CodeGJG
Date Received2007-09-17
Model NumberNA
Catalog NumberMTS9640
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key896101
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-09-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.