DISP FIRSTPASS STR PASSR SELF 22-4038

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,o report with the FDA on 2019-10-15 for DISP FIRSTPASS STR PASSR SELF 22-4038 manufactured by Arthrocare Corp..

Event Text Entries

[162437931] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[162437932] It was reported that during the surgery the jaw of the firstpass jammed shut inside the shoulder while trying to pass tape through the tendon. The black handle of the jaw, including the ratchet, was stuck all the way in and would not release. Eventually, by pulling the black handle with significant force the jaw was released and the device removed from the patient. The needle was inspected as intact. The device break inside the patient and the pieces were retrieved. A backup device was available to complete the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006524618-2019-00515
MDR Report Key9195148
Report SourceFOREIGN,HEALTH PROFESSIONAL,O
Date Received2019-10-15
Date of Report2019-10-15
Date of Event2019-09-24
Date Mfgr Received2019-09-24
Device Manufacturer Date2018-08-20
Date Added to Maude2019-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactHOLLY TOPPING
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer Phone5123913905
Manufacturer G1ARTHROCARE CORP.
Manufacturer Street7000 W. WILLIAM CANNON
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal Code78735
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISP FIRSTPASS STR PASSR SELF
Generic NamePASSER
Product CodeHWQ
Date Received2019-10-15
Catalog Number22-4038
Lot Number2018932
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerARTHROCARE CORP.
Manufacturer Address7000 W. WILLIAM CANNON AUSTIN TX 78735 US 78735


Patients

Patient NumberTreatmentOutcomeDate
10 2019-10-15

© 2026 FDA.report