REP REPORTED THAT THE PERFORATOR DID NOT STOP WHEN GOT BEYOND THE CRANIAL BONE, WHICH RESULTED IN TORN DURA.
| Seq | Brand | Generic | Manufacturer | Product code | Model | Catalog | Lot | PMA | 510(k) | Implant | Evaluated | Availability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | CODMAN DISPOS PERFORATOR | DRILLS, BURRS, TREPHINES, & ACC. | CODMAN & SHURTLEFF, INC. | KAT | NA | 26-1221 | UNK | N | N | R |
| Sequence | Received | Treatment | Outcome |
|---|---|---|---|
| 1 | 2007-10-09 | 0 | 1. R |
Patient 1
REP REPORTED THAT THE PERFORATOR DID NOT STOP WHEN GOT BEYOND THE CRANIAL BONE, WHICH RESULTED IN TORN DURA.
Patient 1
HISTORICALLY FOR COMPLAINTS OF THIS NATURE, THE RETURNED PERFORATORS HAVE BEEN VISUALLY INSPECTED AS RECEIVED, DISASSEMBLED AND CLEANED, AND THEN VISUALLY AND DIMENSIONALLY INSPECTED. NO DISCREPANCIES HAVE BEEN FOUND. THE PERFORATORS HAVE BEEN REASSEMBLED AND WERE FUNCTIONALLY TESTED FOR CUTTING AND DRILLING. THEY'VE BEEN FOUND TO MEET SPECIFICATION REQUIREMETNS. THE REPORTED CONDITION COULD NOT BE DUPLICATED. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE FOUND NO DISCREPANCIES. THE COMPLAINTS HAVE NOT BEEN CONFIRMED. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.