REPROCESSED ETHICON TX60B OPEN LINEAR STAPLER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2007-10-15 for REPROCESSED ETHICON TX60B OPEN LINEAR STAPLER manufactured by Sterilmed, Inc..

Event Text Entries

[706761] User had great difficulty loading the reload cartridge into the device. Once loaded the device misfired into the bowl after several attempts to fire with no results.
Patient Sequence No: 1, Text Type: D, B5


[8120566] The investigation yielded two possible causes that may have led to the issue. One possibility is a problem with the reload cartridge itself. There may have been a flaw in the cartridge making it difficult to load and subsequently misfire. Sterilmed does not supply reload cartridges with reprocessed open linear staplers. Another possibility lies within the device. When the release button is pushed the closing handle trigger should return to its normal position. This action is designed to bring the lower jaw of the device down into the correct position to accept a reload. If the lower jaw is not completely into position a reload can be forcibly loaded into the device, however it will not be seated properly. If this is the case the retaining pin will not seat properly into the anvil and can result in improperly formed staples. During the sterilmed investigation the jaws returned to their normal position without incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2134070-2007-00014
MDR Report Key928756
Report Source99
Date Received2007-10-15
Date of Report2007-10-15
Date of Event2007-09-20
Date Mfgr Received2007-09-20
Device Manufacturer Date2007-01-16
Date Added to Maude2008-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDENNIS TOUSSAINT
Manufacturer Street11400 73RD AVE. NORTH
Manufacturer CityMAPLE GROVE MN 55369
Manufacturer CountryUS
Manufacturer Postal55369
Manufacturer Phone7634883410
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREPROCESSED ETHICON TX60B OPEN LINEAR STAPLER
Generic NameREPROCESSED OPEN LINEAR STAPLER
Product CodeNLL
Date Received2007-10-15
Returned To Mfg2007-09-28
Model NumberTX60B
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key940793
ManufacturerSTERILMED, INC.
Manufacturer AddressMAPLE GROVE MN 55369 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-10-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.